AM-5100 SO2 analyzer adopts ultraviolet fluorescence technique and combines advanced micro processing technology, which could provide accurate and reliable SO2 detection at low level ppb~ppm.This analyzer can be applied to atmospheric monitoring and dilution-based CEMS system.

Features Of AM-5100 SO2 Analyzer

Measurement range 0~500ppb to 0~20ppm, user selectable;

High efficiency hydrocarbon remover and unique optical design, strong anti-interference ability;

Dark current compensation at fixed time, high stability;

Self checking function of light intensity attenuation;

Adaptive signal filtering technology to optimize response time;

Large capacity memory, automatic storage of historical data.

Application Of AM-5100 SO2 Analyzer

Atmospheric monitoring

Specification Of AM-5100 SO2 Analyzer

Range:

0~500ppb, wide range 0~20,000ppb (optional)

Unit:

ppb, ppm, ug/m3, mg/m3

Zero Point Noise:

<0.2ppb (RMS)

Low Detection Limit:

0.4ppb

Zero Drift:

<1ppb/24h

Span Drift:

<5ppb/24h

Response Time:

<120s (T90)

Linearity:

<1%/F.S.

Repeatability:

<1%

Sampling Flow Range:

800SCCM±10%

Ambient Temp.:

5~40

Dimension:

178mm×430mm×635mm

Weight:

<15kg, external pump 5kg

Power:

200V~240V, 50/60Hz, 400W

Analog Output:

10V, 5V, 1V, 0.1V (optional)

Comm.:

RS232, RS485